Preparation step three - Zero Calibrate the SidePak

  • With the Zero Filter attached,
    Press ENTER again.
  • Wait until zero calibration is finished, this will take 60 seconds. The display will countdown.

 

  • With the Zero Filter still attached verify that the reading stays near 0.000 mg/m 3 or below 0.003 mg/m3. If the reading is not near 0.00 mg/m3 then repeat the zero calibration until it is.
  • You may need to press the page key one or more times to return to survey mode.

  • Remove the Zero Filter.